(image for) Midmark M9 AIR VENT BELLOWS ASSEMBLY (DIAPHRAGM STYLE)

Midmark M9 AIR VENT BELLOWS ASSEMBLY (DIAPHRAGM STYLE)

Model: RCB089-1223

$179.97

Description

Midmark M9 High Quality Guaranteed Replacement Parts

Part #RCB089-1223
Midmark Part #002-0375-00
Old Midmark Part #014-0193-00

AIR VENT BELLOWS ASSEMBLY (DIAPHRAGM STYLE)

  • Bronze diaphragm
  • 1/4" compression fittings- both ends
  • Mounts to tubing between temperature gauge and reservoir
  • Activates within 12°F of saturated steam

Fits: Reservoir Tubing
Model This Part Fits: Midmark M9 Autoclaves

Now You Can Upgrade Your Midmark M9 From The Air Vent Bellows Assembly To the Newer Style, Longer Lasting Steam Trap Assembly With This Easy To Install Steam Trap Upgrade Kit Part #MIK078-1207

The Bellows Style With It's Extra Convolutions, Is Designed To Last Longer Than The Diaphragm Style

 


SERVICE TIP: AllClaveParts now offers two different Air Vent Bellows Assemblies: Part #RCB100-1224 and RCB089-1223. There are two different styles of assemblies: the diaphragm style and the bellows style. Here is the difference:

Currently, the RCB089-1223 and Midmark's are both diaphragm style assemblies. This style has only one convolution doing all the work.

The other style is a bellows assembly, the RCB100-1224. This style assembly, offered by AllClaveParts, has several convolutions doing all of the work so it lasts much longer than the diaphragm style. Now you have a choice its the AllClaveParts Advantage!

How To Replace The Midmark M9 Bellows Assembly


Remove The Old Midmark M9 Bellows Assembly

  1. Remove the back panel ( Click Here For Instructions)
  2. Disconnect tube (1) from tee (2)
  3. Pull downward on bellows assembly (3) to remove bellows assembly from condensing tank (4)
  4. Loosen and remove tubes (1 and 5) from bellows assembly (3)

Note: The drawing below shows the old-style bellows. The procedure for the new-style is the same

 

Midmark M9 Bellows Assembly


Install The New Midmark M9 Bellows Assembly

  1. Connect tubes (1 and 5) to bellows assembly (3)
  2. Position bellows assembly (3) and insert tube (5) into tube hole of condensing tank (4)
  3. Connect tube (1) to tee (2)
  4. Install back panel ( Click Here For Instructions)
Fast Delivery

We usually ship the same day! All packages are shipped w/Delivery Confirmation.

100% Secure Payment

Your transaction is secure thanks to the PayPal payment system: SSL (Secure Socket Layer).

Shopper Award
(image for) AllClaveParts Payment Methods
allclaveparts.com widget logo