Description
A Return Box Will Be Shipped First: We must receive the old board before we can ship your replacement
Part #MIB130-4219
Midmark Part #002-0434-03
M9 Control (PC) Board (Refurbished)
- High Quality and Fully Guaranteed!
- PC board performance verified, calibrated and tested (Quality Control test printout included with each board)
- 90 Day Warranty
- See NOTES below for Serial Number information
- Includes The Best Guarantee Anywhere
- This PC Board is Not Returnable Nor Refundable - Warranty Service Only
- Also Includes The Best Technical Support In The Industry
Fits: Chassis mounting plate
Model This Part Fits: Midmark M9 Autoclave
Notes: Part only fits M9 -001 thru -019 with Serial Number Prefixes CZ, DA, DB, DX, DY, FD & OM
How It Works
When we receive your order, we will immediately send you a return box with a prepaid UPS Shipping label for your Old Board to be returned to us.
We must receive the old board before we can ship your replacement
Once we receive your Old Board, we will then ship your Newly Refurbished Board to you.
See Below For Things To Check Before Replacing The Midmark M9 PC Control Board
Click Here For Installation Instructions
Things To Check Before You Replace Your Midmark M9 PC Control Board
There are just a few things you need to check before you order your new PC Board
- Check the Pulse Solenoid (Part #MIS044-1214) Ohms. If the Ohms are much lower than ~14.6 on the M9 (Blue coil) or ~11.6 Ohms on the M9(Black coil) then the Pulse Solenoid is bad and will damage the PC Board
- Check for a bad ground connection at the PC Board or a missing/loose screw that is used to mount the PC Board to the mounting plate. Either of these conditions can cause the fill solenoid valve to continually fill the chamber with water thus staying in the fill mode and not going into the sterilize mode
- Check the Water Level Sensor Part #MIS075-1217. A bad Water Level Sensor will do the same thing as a bad ground connection. Keep in mind that if the sensor is not tight, the insulated inserts can take on water and fail to insulate thus causing the PC Board to "think" that the chamber is full of water. In this case, the sterilizer will overheat resulting in a 001 Error Code
- Check the Heater Element Part #MIH048-1198. The M9 Heater elements should read ~10 Ohms. And from ground to one tab of the heater should read OPEN. If the heater is shorted, it will cause damage to the PC Board
Written By Jim Wisniewski Manager, RPI Product Development