(image for) Midmark M9 PULSE SOLENOID

Midmark M9 PULSE SOLENOID

Model: MIS044-1214

$92.97

Description

 

High Quality Midmark M9 Autoclave Repair Parts

Part #MIS044-1214
Midmark Part #002-0363-00

PULSE SOLENOID

  • 3/16" quick disconnect terminals
  • Includes all parts as shown
  • Single would coil 120V/60Hz.
  • Stroke: 1/2"; 14.6 Ω
  • Pull type solenoid
  • Blue Coil 14 ohms

The Pulse Solenoid Automatically Opens The Door When The Cycle Has Been Completed

Installation Instructions Below

Fits: Door hinge mounting bracket
Model(s) This Part Fits: Midmark M9 Autoclave

Notes: Fits Midmark M9 Autoclave -001 thru -019 with Serial # prefixes CZ, DA, DB, DX, DY, FD & OM

Where is the Midmark M9 Pulse Solenoid located?

Look for it in the picture below - It is on the right hand side (as you are facing it) and towards the front of the autoclave

 

Here is where the Midmark M9 Pulse Solenoid is located
 
How To Replace The Midmark M9 Pulse Solenoid

How To Remove The Old Midmark M9 Pulse Solenoid

  1. Remove The Control PC board
  2. Remove two screws (1) from Pulse Solenoid (2)
  3. Remove nut (3) and shoulder screw (4), and then remove pulse solenoid (2) and latch lever (5) as an assembly from pressure vessel bracket (6)
  4. Using a 3/32 in. punch (available at any hardware store), remove roll pin (7) and latch lever (5) from plunger of pulse solenoid (2)
Midmark M9 Pulse Solenoid


How To Install The New Midmark M9 Pulse Solenoid

  1. Using a 3/32 in. punch, install latch lever (5) on plunger of pulse solenoid (2) and secure using roll pin (7)
  2. Coat shoulder screw (4) with Loctite 271
  3. Install latch lever (5) and pulse solenoid (2) as an assembly on pressure vessel bracket (6)
  4. Secure using shoulder screw (4) and nut (3)
  5. NOTE: The holes for the two screws (1) are oblong. Install screws at the top of each hole first, then adjust downward as necessary if door will not open.
  6. Position pulse solenoid (2) on chamber bracket (6) and secure in position using two screws (1)
  7. Install control PC board
"Midmark M9 Autoclave Repair Made Simple™"

 

Fast Delivery

We usually ship the same day! All packages are shipped w/Delivery Confirmation.

100% Secure Payment

Your transaction is secure thanks to the PayPal payment system: SSL (Secure Socket Layer).

Shopper Award
(image for) AllClaveParts Payment Methods
allclaveparts.com widget logo