Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Parts
›
Midmark ® Ritter
›
Midmark M9 & M9D
› How To Replace The M9 Fill Solenoid Valve
How To Replace The M9 Fill Solenoid Valve
Midmark M9 & M9D Replacement Parts & Installation Guides
PART # MISO42-1212
OEM PART # 002-0365-00
Midmark M9 Fill Solenoid Valve Removal / Installation Instructions
NOTE
Steps will vary slightly for M9D units. Use this procedure as a guide.
Removal
NOTE
Units prior to serial number CZ1110/OM1000 have a manifold assembly which secures the fill solenoid.
Refer to the parts list to remove the fill solenoid on these sterilizers.
Drain water from condensing tank.
Remove back panel (
See Instructions Here
).
Tag and disconnect two wires (1, Figure Below) from terminals (3).
Tag and disconnect two wires (2) from terminals (4).
Disconnect three tubes (5 thru 7) from elbows (8 thru 10).
Remove two screws (11) from bottom of manifold assembly (12).
NOTE:
Spacer (12A) is only on newer units.
Remove manifold assembly (12) and spacer (12A) from base (13).
Hold screw (1, Figure 4-13) and loosen nut (2). Turn terminals out of the way.
Remove assembled vent solenoid (3) and elbow (4) from tee (5).
Remove elbow (6) and tee (5) from fill solenoid (7).
Remove elbow (8) from fill solenoid (7)
Installation
Coat threads of tee (5, Figure Below) and elbows (6 and 8) with teflon tape.
Install elbow (8) on fill solenoid (7).
Install elbow (6) and tee (5) on fill solenoid (7).
Loosen nut (9). Turn terminals of fill solenoid (7) so wires will connect easily to fill solenoid. Tighten nut (9).
Coat threads of elbow (4) with teflon tape.
Hold screw (1) and loosen nut (2). Turn terminals out of the way.
Install assembled elbow (4) and vent solenoid (3) on tee (5).
Turn terminals of vent solenoid (3) so wires will connect easily to vent solenoid.
Hold screw (1) and tighten nut (2).
Warning: When installing the vent solenoid and the fill solenoid, the side of the solenoids marked "IN" need to be connected to the pressure lines. Reversing solenoids will cause system failure.
Install spacer (12A, Figure 4-12) and manifold assembly (12) on base (13) and secure using two screws (11).
Connect three tubes (5 thru 7) to elbows (8 thru 10).
Connect two wires (2) to terminals (4).
Connect two wires (1) to terminals (3).
Install back panel (Refer to para 4.5).
Refill condensing tank with distilled water.