(image for) OCR PRESSURE RELIEF VALVE

OCR PRESSURE RELIEF VALVE

Model: PCV052-144

$96.89

Description

Guaranteed Quality Pelton Crane OCM Autoclave Repair Parts

Part #PCV052-144
Pelton Crane Part #004146 / 002-0359-00

PRESSURE RELIEF VALVE

  • 1/2 MPT x 1/2 FPT
  • Cracking Pressure: 35 PSI

Fits: Chamber
Model This Part Fits: Pelton Crane OCR Autoclave

If you see evidence the Pelton Crane OCR Autoclave Pressure Relief Valve is Leaking, Replace it right away.

Even a small drip will act like a leak and keep the autoclave from reaching the proper temperature and pressures needed for sterilization. If this happens, your instruments Are Not Being Sterilized. So replace it immediately

Tech Tip: To remove the Pressure Relieve Valve from the chamber of the Pelton Crane OCR Autoclave, simply turn it in a counter-clockwise. We use a pipe wrench to remove them. It makes the job a whole lot easier!

 

 

HOW TO REPLACE THE PELTON CRANE OCR AUTOCLAVE SAFETY VALVE

 

 

 

 

The Pressure Relief Valve on The Pelt0n Crane OCM & OCR Autoclaves is located towards the rear and top of the chamber

 

 

  1. Unplug Autoclave And Ensure No Pressure Is In The Chamber
  2. Remove The Outer Casing
  3. Locate The Valve at Upper Left Rear Of Camber
  4. Remove The Valve as outlined in the Tech Tip Above
  5. Install the New Valve (Using Plumber's Putty on Threads If Needed To Assure A Tight Seal
  6. Re-use the Directional Pipe Attached To The Old Valve
  7. Manually Activate Valve By Pulling Up On The Ring Periodically to Check Proper Function & Seating
  8. Re-Install Outer Casing

 

 

"Pelton Crane OCR Autoclave Repair Made Simple™"

 

 

Fast Delivery

We usually ship the same day! All packages are shipped w/Delivery Confirmation.

100% Secure Payment

Your transaction is secure thanks to the PayPal payment system: SSL (Secure Socket Layer).

Shopper Award
(image for) AllClaveParts Payment Methods

Customers who bought this product also purchased...

allclaveparts.com widget logo