Guaranteed Quality Pelton Crane Autoclave Repair Parts

Pelton Crane Door Gasket Cross-Section View Part #PCG066-607
Pelton Crane Part #1539407
Old Pelton Crane Part #019763

DOOR GASKET

 

  • OD: 11.312"
  • When replacing this gasket, it is also a good time to replace the Door Spacer ( Part #PCS067-621 )

Model This Part Fits: Pelton Crane Validator 10 Autoclaves

Order the Sterilizer PM Kit: It includes this part and more! Part #PCK221-616

 

Fits: Door

Pelton Crane Validator 10 Door Gasket Installation Procedure

Remove the old door gasket. Clean the gasket groove and chamber face ring surface.

For easy install of the new gasket, use the "star" or "North-South-East-West" method (ie. begin installing the gasket at the top of the door, then the bottom, then the left side, then the right side, and then eveything in between making sure that the gasket fits flush without bumps or gaps).


 