Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Parts
›
Ritter - Castle
›
777 Parts Installation Guides
› Installing The Castle 777 Door Gasket
Installing The Castle 777 Door Gasket
The Castle 777 Autoclave Is A Classic And If You Are Fortunate Enough To Own One, You Have A Prized Possession!
How To Replace The Castle 777 Door Gasket
For Easy Installation Of The Door Gasket, Just Follow These Instructions
Remove the Old Gasket
Clean The Door Channel & Make Sure It Is Clear Of Any Foreign Matter Or Debris, And the edge of the lip is not dented or marred.
Make Sure Door Channel Is Clean Before Proceeding
Place New Door Gasket In Position By Squeezing It In As Tight As Possible With Your Fingers
Make Sure Gasket Is Evenly Distributed All Around Opening
Prepare Atuoclave For Regular Operation (Pour In Water)
Close Door Carefully, Making Sure Gasket Remains In Place
Turn the Autoclave On and Run It Up To 15 psi. The Gasket Should Squeeze Into Place At This Pressure
Allow The Unit To Run For 15 Minutes at 15 psi. Then Turn It Off & Vent Steam
After The Unit Has Cooled Down, Inspect The Door Gasket To Be Sure It Is Equally Inserted At All Places
If The Gasket Can Be Rolled Out By Rubbing Your Thumb From Front To Back Over The Lip Of The Door, It Is Not Properly Placed
If The Placement Is Not Proper, Repeat Operation Of The Autoclave For A Second Cycle
Note:
After Installing The New Castle 777 Door Gasket, The Door May Leak At Low Pressure for Several Cycles Until It Seats Itself.