Installing The Castle 777 Door Gasket

We carry an extensive line of high quality autoclave repair parts. Plus, every part you purchase from us, comes with the best technical support in the business





The Castle 777 Autoclave Is A Classic And If You Are Fortunate Enough To Own One, You Have A Prized Possession!



How To Replace The Castle 777 Door Gasket



For Easy Installation Of The Door Gasket, Just Follow These Instructions
  1. Remove the Old Gasket
  2. Clean The Door Channel & Make Sure It Is Clear Of Any Foreign Matter Or Debris, And the edge of the lip is not dented or marred. Make Sure Door Channel Is Clean Before Proceeding
  3. Place New Door Gasket In Position By Squeezing It In As Tight As Possible With Your Fingers
  4. Make Sure Gasket Is Evenly Distributed All Around Opening
  5. Prepare Atuoclave For Regular Operation (Pour In Water)
  6. Close Door Carefully, Making Sure Gasket Remains In Place
  7. Turn the Autoclave On and Run It Up To 15 psi. The Gasket Should Squeeze Into Place At This Pressure
  8. Allow The Unit To Run For 15 Minutes at 15 psi. Then Turn It Off & Vent Steam
  9. After The Unit Has Cooled Down, Inspect The Door Gasket To Be Sure It Is Equally Inserted At All Places
    If The Gasket Can Be Rolled Out By Rubbing Your Thumb From Front To Back Over The Lip Of The Door, It Is Not Properly Placed
  10. If The Placement Is Not Proper, Repeat Operation Of The Autoclave For A Second Cycle
Note: After Installing The New Castle 777 Door Gasket, The Door May Leak At Low Pressure for Several Cycles Until It Seats Itself.