Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Parts
›
Scican Statim
›
Statim 5000 Parts Install Guides
› Statim 5000 Thermocouple Installation Guide
Statim 5000 Thermocouple Installation Guide
PART #SCT030-4017
Statim 5000 THERMOCOUPLE
INSTALLATION INSTRUCTIONS
Disconnect the power to the sterilizer.Remove the cover. Disconnect the two cables from the PC Board - one cable from the LCD Display Board and the other from the Keypad
Inspect the old thermocouples tip position from the inside of the cassette bay (armature) for later reference. See Figure 1.
If replacing the thermocouple because of a bent tip, straighten to allow for removal, see Service Tip.
Disconnect the three (3) thermocouple wires at the PCB (see Figure 2) and cut all securing cable ties.
At the rear of the probe bracket, locate the old thermocouple fitting and loosen the large inner compression nut, then slide out the old thermocouple with the small compression fitting assembly (see Figure 4).
Measure the old thermocouple from the tip to the end of the compression fitting assembly as shown in Figure 4.
Use the measurement from Step #4, apply that same measurement to the new thermocouple.
NOTE:
For proper function, it is very important that the same measurement from the old thermocouple is applied to the new thermocouple so that the position of the new assembly when installed is the same as the old assembly.
Hand tighten the small compression fitting nut down onto the new thermocouple and re-check the measurement.
Using two (2) 5/16 wrenches, tighten down the small compression nut.
CAUTION: Once the fittings are secured with the wrenches, the position cannot be changed.
Now make a pencilmark on the top of the small compression fitting nut (see Figure 5).
Pre-bend thermocouple tip as shown. Insert the new thermocouple assembly back into the probe bracket assembly and hand tighten the large compression fitting nut (see Figure 4).
Be sure to keep the bend upwards by ensuring the mark on the small compression fitting nut is up (see Figure 4 & 5).
Look inside the cassette bay (armature), verify that the tip of the new thermocouple is positioned slightly upwards as shown in Figure 1.
Tighten the large compression nut using one 5/16" wrench and one 1/2" wrench.
Carefully insert the cassette to verify proper fit.
Bend the outer part of the new thermocouple up and over the probe bracket.
CAUTION: Do not make sharp bends or kink the new thermocouple tube, maintain a minimum bend radius of 3/16 (5mm). See Figure 4.
Run the wiring back to the PCB, secure using the Cable Ties (Part #RPT083) provided in this package. Refer to Figure 3.
Reconnect the ribbon cables from the cover to the PCB (for ease of operation use an Extension Test Cable
(Part #SCT026-527)
. See Figure 3.
Run an unwrapped cycle and check for steam leaks at all new and existing connections. If leaks are found, repair themand repeat testing as necessary.
If attached, remove Extension Test Cable and replace the cover. Repeat test cycle.