Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Parts
›
Tuttnauer ®
›
1730 Series
›
1730 Parts Installation Guides
› Front Leveling Feet Installation Instructions
Front Leveling Feet Installation Instructions
Tuttanuer 1730 Series FRONT FEET
ADJUSTMENT & INSTALLATION INSTRUCTIONS
After replacing the Front Feet
It is recommend that you check the chamber pitch to ensure proper water fill - Refer to Chamber Pitch Adjustment below for procedure.
If the Water Fill Elecrtrode needs to be readjusted, refer to Adjustment of the Water Fill Electrode below for procedure.
If servicing any "E" Series models, refer to Checking and Adjusting the Water Fill Time to the right for final testing procedure.
Chamber Pitch Adjustment
Adjusting the chamber for proper pitch is essential to the performance of the sterilizer. If the pitch is not set correctly, too much or too little water will fill into the chamber at the beginning of each cycle preventing the sterilizer from performing optimally.
To adjust the pitch of the chamber:
Empty the sterilizer.
Turn
OFF
the sterilizer.
Important:
The chamber must be dry and cool to the touch before proceeding to the next step.
Ensure that all legs are installed properly and the unit is located on a sturdy, level surface.
Measure 10 ounces (300 ml) distilled water and pour through the open door directly into the chamber. The water must reach the indication groove near the front the chamber. Refer to Figure 1 below.
If the water does not reach or goes past the indication groove, the sterilizer must be adjusted. To help level the sterilizer, the front legs of the sterilizer may be adjusted using a wrench
Once the chamber pitch has been properly set, drain and dry the chamber
Adjustment of the Water Fill Electrode
Unplug the sterilizer.
Ensure that the chamber is cool and dry.
Remove the back panel of the unit.
Locate the Water Fill Electrode which is mounted in the bottom of the chamber. Carefully push aside the insulation covering the electrical connection.
Loosen the compression nut holding the electrode in place. See Figure 2 below for a complete illustration of the various parts involved and the orientation of the ferrule and the compression nut.
The electrode can be adjusted in and out of the chamber.
Measure out 7 ½ ounces of distilled water and pour it into the chamber ensuring that none of the water leaks out from the compression fitting.
Adjust the Water Fill Electrode by sliding it further into or out of the compression fitting until the tip just breaks the top of the water. See Figure 3 below.
Tighten the compression fitting until the ferrule is seated and properly compressed being careful not to over tighten it. Carefully remove as much of the water as possible from the chamber.
Plug in the unit and run a cycle to check for steam leaks around the compression fitting
Checking and Adjusting the Automatic Water Fill Time
A. This procedure applies to E Series sterilizers that DO NOT have built in water pumps. (The serial number/date code does not contain the letters WP.)
Ensure that the chamber is dry, the reservoir is full, and that the sterilizer is plugged in and turned ON.
Open the chamber door and locate the door switch. Press the door switch in, and hold it. At the same time, press and release the START key. Release the door switch after water starts to flow into the chamber.
The final water level should reach the groove at the front of the chamber. If the water level reaches the groove, no adjustments are necessary. If the water level does not reach the groove, proceed to Steps #4-11 to make adjustments.
Turn OFF the sterilizer. Ensure that the chamber is dry, the reservoir is full and that the sterilizer is plugged in.
Locate the WATER INLET key on the front panel keypad. (Note: It is the key with the two arrows between parallel lines.)
With the door open, press and hold the WATER INLET key and turn power ON.
Release the WATER INLET key when the normal program display screen appears. Wait one second then press and hold the WATER INLET key again.
Water should begin to flow into the chamber.
Release the WATER INLET key when the water level has reached the groove at the front of the chamber.
Wait ten seconds for the system to record the new fill time into memory before pressing any additional keys or turning off the sterilizer.
Empty the water from the chamber, then repeat Steps #2-3 to check water level.
B. This procedure applies to E Series sterilizers that DO have built in water pumps. (The serial number/date code ends with the letters WP.)
Service Tip:
If it becomes necessary to reset the software in an E series model sterilizer with a water pump, this procedure MUST be followed before using the sterilizer to ensure the correct amount of water will be delivered to the chamber.
Ensure that all legs are installed properly, none of them are damaged,and the unit is located on a sturdy, level surface.
Ensure that the chamber is dry, the reservoir is full, the door is closed and properly sealed. The sterilizer must be plugged in and turned ON.
Press the STOP key on the front panel keypad repeatedly until the message CODE: XXX appears.
Using the UP/DN arrow keys, change the displayed message to CODE: 105 then press the STOP key.
Ensure that the message WATER IN = XX SEC is displayed. Using the UP/DN arrow keys, change the seconds displayed to the 30 Seconds.
NOTE:
Fill times are suggested starting values - if during daily use there appears to be inadequate water in the chamber then fill times can be increased to add additional water to the chamber.
Press the STOP key to set the displayed value in memory.
The message EA TYPE: should be displayed at this point.
Using the UP/DN arrow keys, select either â0â for a model E or EK unit, or select "1" for an EA or EKA unit.
Press the STOP key to finish programming the fill time.
Run one or two cycles to check for proper operation.