(image for) 1730EK CUT OUT THERMOSTAT (MANUAL RESET)

1730EK CUT OUT THERMOSTAT (MANUAL RESET)

Model: TUT038-6673

$148.97

Description



Tuttnauer Manual Reset Cut-Out Thermostat Part #TUT038-6673
Tuttnauer Part #01620004

CUT OUT THERMOSTAT (MANUAL RESET)

  • Temperature range: 32°F to 450°F (0°C to 230°C)
  • Preset 450°F (230°C)
  • Stainless steel bulb and capillary tube

Fits: Mounting bracket in rear of machine
Model This Part Fits: Tuttnauer 1730EK Autoclaves

Click Here For Step-By-Step Instructions For Testing Your Tuttnauer Manual Cut Out Thermostat

 

 

Tuttnauer Manual Cut Out Thermostat Location

 

 

 

 

 


This thermostat cuts off power to the autoclave, in the event that all other safety systems fail.

As an example: If the fill cycle is skipped and there is no water in the chamber when the Sterilization Cycle is started, the chamber will heat up and activate the cut-out thermostat.

If the cut-out thermostat is activated, press the RESET button on the back of the unit.

If the autoclave is operated according to the instructions, and the thermostat activates, cutting off the power, replace the thermostat.

The thermostat has been calibrated at the factory. Do not attempt to re-calibrate it

 

 

 