Description



Tutnauer Manual Reset Safety Thermostat Part #TUT038-3301
OEM Part #01620004



SAFETY THERMOSTAT (MANUAL RESET)

 

 

  • Temperature range: 32°F to 450°F (0°C to 230°C)
  • Preset 450°F (230°C)
  • Stainless steel bulb and capillary tube

Fits: Mounting bracket in rear of machine

Click Here For Step-By-Step Instructions For Testing Your Tuttnauer Manual Cut Out Thermostat

 

 

 

 

 

 

 

 

 


This thermostat cuts off power to the autoclave, in the event that all other safety systems fail.

As an example: If the fill cycle is skipped and there is no water in the chamber when the Sterilization Cycle is started, the chamber will heat up and activate the cut-out thermostat.

If the cut-oput thermostat is activated, press the RESET button on the back of the unit.

If the autoclave is operated according to the instructions, and the thermostat activates, cutting off the power, replace the thermostat.

The thermostat has been calibrated at the factory. Do not attempt to re-calibrate it

 

 

 