Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Parts
›
Tuttnauer ®
›
3870 Series
›
Parts Installation Guides
›
Safety Thermostat
› Remove & Replace The Tuttnauer 3870 Series Safety Thermostat
Remove & Replace The Tuttnauer 3870 Series Safety Thermostat
Tuttnauer 3870 Series Safety Thermostat Replacement
(Installation and replacement procedures are identical for both the Automatic Reset Safety Thermostat and the Manual Reset (Cut-Out) Thermostat)
NOTE:
DO NOT ATEMPT TO ADUST THIS THERMOSTAT
This Thermostat is calibrated at the factory and does not require calibration
Tampering with the adjustment on this thermostat will totally defeat the safety feature.
The only recommendation is that a defective thermostat be replaced with a factory calibrated replacement.
Removing The Old Safety Thermostat
Unplug the unit
Remove the wires from the terminals
Tech Tip:
Tag Each Wire As It Is Removed To Simplify Reconnecting To The New Safety Thermostat
Remove the two screws holding the Safety Thermostat to the mounting bracket, or unscrew the lock nut, which ever method is being used to secure the thermostat
Loosen the bolts on the rear most Heating Element or which ever element the probe is held in place by. Remove the probe
Installing The New Safety Thermostat
Carefully uncoil the tubing connecting the probe to the contact body
The probes and the tubing leading to the body of the thermostat are filled with liquid. It is important that while working with these sensors that the tubing not be kinked.
If the tubing is kinked or the probe punctured the device will not work and should be replaced.
Mount the new thermostat body using the screws or lock nut that were removed in step 3 above.
Carefully route the probe and tubing over to the rear Heating Element
The new probe should be installed in the proper channel on the rear most Heating Element, regardless of what element the probe was under originally
The Safety Thermostat is in the lower channel
The Cut-Out Thermostat is in the upper channel
The body of the Cut-Out Thermostat is mounted on the top portion of the mounting bracket.
The body of the Safety Thermostat is mounted on the bottom portion of the mounting bracket.
NOTE:
Improper mounting of the probes will result in a malfunctioning of the autoclave.
If the replacement thermostat has a long probe it is only necessary to secure the rear portion of the probe under the rear most element.
Retighten the Heating Element making sure that the probe is held securely between the element and the Chamber.
It is important that the probes be snug in channels of the Heating Elements. If they are loose they will not make good contact with the Chamber and well not properly sense the temperature.
Flattening the channel on the Heating Element slightly is an acceptable way of snugging the probe.
Reconnect the wires to the same terminals that they were removed them
Be careful not to connect a high voltage wire to the ground terminal.
If the replacement thermostat does not have a separate ground terminal then the ground wire need not be connected. Screwing the thermostat to the mounting bracket will supply sufficient grounding.