Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
AMSCO 3000 BASIC TROUBLESHOOTING
› AMSCO 3000 BASIC TROUBLESHOOTING GUIDE
AMSCO 3000 BASIC TROUBLESHOOTING GUIDE
Amsco Eagle 3000 Basic Troubleshooting For
Amsco Eagle 3000 Gravity & Vacumatic Sterilizers
By AllClaveParts.com
PROBLEM
POSSIBLE CAUSE AND/OR CORRECTION
No Power
Check facility power source turn on.
POWER switch OFF turn ON
Unable To Select Cycle
Chamber door(s) open close and lock door(s). (Control will beep three times if door is open when cycle is selected.)
Chamber Temperature Does Not Rise To Correct Setting
Chamber drain line or strainer plugged clean line or strainer
Chamber Pressure & Temperature Drop During A Cycle
Chamber drain line or strainer plugged clean line or strainer
Pressure Too High, Temperature Too Low During Sterilization Phase
Air trapped in chamber because of clogged strainer or drain clean strainer inside chamber, flush and clean drain line and traps
Steam Escapes Around Door During Sterilization Phase
Replace Worn Or Defective Door Gasket
Wrapped Items Are Wet After Completion Of Cycle, Or Excessive Vapor In Chamber When Door Is Opened
Dry time too short for load type increase setting
Water In Chamber Or Door Channel At End Of Cycle
Drain plugged or clogged clean drain.
Steam separator/strainer plugged clean.
Drain line check valve defective replace.
Jacket Pressure Does Not Rise To Required Level
Steam supply hand valve not open open valve.
Possible flood condition/steam supply solenoid valve closed
Steam supply less than 50 psig (3.52 kg/sq.cm) dynamic check supply pressure & increase if necessary
Cycle does not advance from prevacuum phase
Door gasket worn or defective replace
Chamber drain line or strainer plugged clean line or strainer
Water supply hand valve not open open valve
Water supply less than 30 psig (2.11 kg/sq.cm) dynamic check supply pressure and increase if necessary.
Steam supply insufficient see 9 above; Jacket pressure does not rise . . .
HI-LO valve not set correctly check that it is set on HI.