Symptom |
Cause |
Solution |
Mineral buildup on chamber wall or water spots on chamber and contents |
Minerals in water or on contents (using tap water |
Clean with Omni-Cleaner Plus® following instructions on bottle. Scrub chamber with Bon-Ami® or Scotch-Brite®. Use distilled water only. |
Unit takes an excessive amount of time (over 20 minutes from a warm start)to reach required sterilization temperature. |
- Load too large. Steam pushing out air too slowly
- Leak in valves.
|
- Remove some packages/articles for faster steam flow
- Call authorized service representative
|
Moisture evident in packs or wraps after drying cycle |
- Overloaded chamber.
- Chamber tilted forward resulting in water left in chamber after venting
- Fill line filter (inside chamber) clogged or not properly installed, resulting in too much water remaining in chamber after venting.
- Drying time not adequate for load.
- Drying air intake filter clogged.
- Drying pump not operating properly
- Unit run without preheating
|
- Follow “Preparation and Loading” instructions
- Level unit properly. Click Here for the proper leveling procedure of your Delta Q
- Clean filter in ultrasonic cleaner. Adjust fill line filter to center bottom of chamber. Replace if air and water cannot freely pass through filter.
- Increase drying time. (see Programming)
- Call authorized service representative
- Call authorized service representative
- Wait 30 minutes to start sterilizing cycle after power is turned on.
|
Wet layer in pack |
Inadequate drying air penetration due to tight pack wrapping or condensation from packs |
Inspect pack for proper wrap. Basins, kits and other utensils should be prepared as individual packs. |
Water collects in bottom of chamber when unit is not in “Fill” cycle. |
- Leaking fill/solenoid valve
- Hole in condensing coil
|
- Rebuild Fill Valve
- Replace Condensing Coil
|
Unit drips on counter from door during cycle |
- Debris on gasket or door ring
- Cut in gasket or excessive wear
|
- Clean door gasket and door ring
- Replace door gasket if cut or worn.
|
Water comes out of fill opening on top of case when unit vents to reservoir. |
- Reservoir overfilled
- Cracked condenser tube causing turbulence under fill opening.
|
- Drain some water. Fill to bottom of openings in fill cup.
- Replace condenser tube.
|
Unit will not turn on |
- No power to wall receptacle
- Master power switch is not turned on
- Defective electronics
|
- Reset building circuit breaker
- Be sure main power switch/circuit breaker on back of unit is on
- Contact us
|
Dark stains on instruments |
- Electroplating
- Cold sterilization residue
- Chlorine stains
|
- Do not mix dissimilar metals in the same package. Separate carbon steel, aluminum chrome and brass from stainless steel trays by using a tray liner
- Wash thoroughly before loading
- Do not use toweling or packaging which may contain chlorine bleach residue
|
Chamber surfaces become cracked |
Chlorine |
NEVER use chlorine cleansers or materials in chamber. Wrapping materials should not be cleaned with chlorinated bleaches. Use only distilled water. |
Unit clicks, but no display; unit shuts off intermittently in mid-cycle |
Malfunctioning connector cable |
Contact us |
Door leaks on Delta Q10 autoclave |
Lip of tray rest is in front of the water dam |
Place lip behind water dam. |
Instruments become rusted or stained |
Minerals and metals in reservoir or chamber |
Use only distilled or deionized water for autoclaving. Clean and flush reservoir/chamber. |
Positive biological indicator (Fails Spore Test) |
- Insuffcient steam penetration
- Wrong type of biological indicator for this sterilizer, wrong cycle used or biological indicator has expired
|
- Reduce load weight or check if load is packed properly.
- Check bioligical indicator instructions to make sure indicator is for a gravity type sterilizer. Check that indicator is correct; one for cycle times and temperatures being run. If necessary, contact your biological indicator representative.
|