ERROR CODE C060: POWER MODE

Free Autoclave Troubleshooting Guides



Midmark M9 & M11 Error Code C060: POWER MODE


Display/Symptom Cause What to Check Solution
C060: POWER MODE

SYSTEM HARDWARE

ITEMS NOT STERILIE
UNPLUG/RE-PLUG UNIT
Power interruption

Previous Error Code may not have been properly cleared before power was connected.
Check power cord connnections

Check voltage supply to assure outlet is on a dedicated, properly rated circuit.
Unplug unit for 60 seconds, then plug unit into a dedicated, properly rated circuit

(Power interruption could be due to an electrical storm, brown out, etc.)
Loose / damaged wire connection(s) Check wire connections at these locations:
  • J5 on main PC board
  • AC receptacle terminals
Secure, repair or replace loose or damaged connections

allclaveparts.com widget logo