Display/Symptom |
Cause |
What to Check |
Solution |
C232: FILL MODE
WATER LOW
ITEMS NOT STERILIE
PUSH STOP TO RESTART
Water level sensor did not detect water within the 5 minute time limit
|
Reservoir water level is low. |
Check water level in reservoir |
Add water to the reservoir until level is in the correct range
Press STOP button, then initiate a new cycle
Press STOP button, then initiate a new cycle
|
|
Chamber filter is dirty or clogged |
Inspect the Chamber Filter |
Clean or Replace the Chamber Filter
Click Here for the Replacement Chamber Filter For The Midmark M9 Autoclave (Part #MIF062-1195)
Click Here for the Replacement Chamber Filter For The Midmark M11 Autoclave (Part #MIF062-798)
|
|
Water Level Sensor is dirty or defective |
Inspect Water Level Sensor |
Clean or Replace the Water Level Sensor
Click Here for the Midmark M9 Water Level Sensory Assembly (Part #MIS075-1217)
Click Here for the Midmark M9 Water Level Sensory Assembly (Part #MIS075-815)
|
|
Loose / damaged wire connection(s) |
Check wire connections at these locations:
- J8 on main PC board
- Water level sensor terminal
|
Secure, repair or replace loose or damaged connections |
|
Chamber is dirty or corroded |
Inspect the bottom of the chamber |
Clean the bottom of the Chamber with Mid-Clean™ and distilled water Part #MIC143-6136 |
|
Fill Valve or PC Board malfunctioning |
Inspect Fill Valve and related tubing for blockages |
Clean, repair or replace Fill Valve |
|
|
Inspect two wires connecting the Fill Valve and PC board (W3 & W4 on PC board) for damage and/or loose connections |
Secure any loose connections |
|
|
With the sterilizer in the Fill Mode...Perform: Fill Valve Testing |
Follow instructions in test procedures |
|
Air valve or PC board malfunctioning |
With the sterilizer in the Fill Mode, Perform Air Valve Testing |
Follow instructions in test procedures |