ERROR CODE C382: FILL MODE

Free Autoclave Troubleshooting Guides






Midmark M9 & M11 Error C382 (FILL)





Display/Symptom Cause What to Check Solution
C382 FILL MODE
DOOR OPEN
ITEMS NOT STERILE
CLOSE DOOR
Door switch contacts opened during Fill Mode Check with operator to determine if door was opened during Fill Mode Explain proper operation
Close door
Cycle will continue where it left off
If the door is hard to close, inspect door & dam gaskets for proper installation

Be sure trays are pushed all the way in
Install door / dam gaskets properly

Remove any obstructions that may prevent the trays from sliding in completely
Perform: Door Switch Continuity Test Follow instructions in test procedures