Midmark M9 & M11 C383 (HEATUP)
Display/Symptom |
Cause |
What to Check |
Solution |
C383 HEATUP MODE
DOOR OPEN
ITEMS NOT STERILE
CLOSE DOOR
|
Door switch contacts opened during Fill Mode |
Check with operator to determine if door was opened during Fill Mode |
Explain proper operation
Close door
Cycle will continue where it left off |
|
|
If the door is hard to close, inspect door & dam gaskets for proper installation
Be sure trays are pushed all the way in |
Install door / dam gaskets properly
Remove any obstructions that may prevent the trays from sliding in completely |
|
|
Perform: Door Switch Continuity Test |
Follow instructions in test procedures
Click Here for the Midmark M9 Door Switch Part #MIS095-4036
Click Here for the Midmark M11 Door Switch Part #MIS095-4035 |