Midmark M9 & M11 Error Code C534: STERILIZE MODE
Display/Symptom |
Cause |
What to Check |
Solution |
C534: STERILIZE MODE
STEAM TEMP HIGH
ITEMS NOT STERILE
UNPLUG / REPLUG UNIT
During Sterilization Mode, chamber temperature reached 6°F above the designated sterilization temperature |
Pressure leaking from chamber |
Check for pressure leaks. Click Here To See "How To Find Pressure Leaks" |
Repair or replace malfunctioning component |
|
Temperature Sensor malfunctioning |
Be sure there is nothing touching the temperature sensor. (Example: trays, pouches, cassettes, etc.) |
Reposition items so nothing touches the temperature sensor |
|
Sterilizer is not level |
Check surface where autoclave is located to make sure it is level |
Level with From Leveling Feet, or move the autoclave to a level surface |
|
Heating element malfunctioning |
Test the Heating Element |
Replace The Heating Element if it fails the test
Click Here and scroll down the page for the instructions on testing the Midmark M9 autoclave
Click Here and scroll down the page for the instructions on testing the Midmark M11 autoclave
|