ERROR CODE C560: POWERUP MODE

Free Autoclave Troubleshooting Guides






Midmark M9 & M11 Error Code C560: POWERUP MODE


Note: When an error code ends in "0", it means the unit had a power interruption during operating conditions. The operator may have unplugged the unit when a previous error code was displayed

Display/Symptom Cause What to Check Solution
C560: POWERUP MODE
STEAM TEMP HARDWARE
UNPLUG / REPLUG UNIT

Steam A/D converter reports an average value outside the limits for normal operation during POWER UP mode.
Unit was brought in from cold environment. Allow unit to warm to room temperature Unplug unit for 60 seconds, then plug unit into a dedicated, properly rated circuit
Temperature Sensor malfunctioning Check temperature sensor harness connection at J12 on PC board Secure loose and/or damaged connection
Perform: Temperature Sensor Testing Follow instructions in test procedures.