Midmark M9 & M11 Error Code C562: HEATUP
Display/Symptom |
Cause |
What to Check |
Solution |
C562: HEATUP
STEAM TEMP HARDWARE
UNPLUG / REPLUG UNIT
Steam A/D converter reports an average value outside the limits for normal operation during POWER UP mode. |
Temperature Sensor malfunctioning |
Be sure there is nothing touching the temperature sensor. (Example: trays, pouches, cassettes, etc.) |
Reposition items so nothing touches the temperature sensor |
|
|
Check the temperature sensor probe for corrosion or residue. |
Clean the temperature sensor with Mid-Clean and distilled water. Click here to get Mid-Clean (Part #MIC143-6907) |
|
|
Check temperature sensor harness connection at J12 on PC board |
Secure loose or damaged connection |
|
|
Perform Temperature Sensor Testing |
Follow instructions in test procedures |
|
Sterilizer is not level |
Check surface where autoclave is located to make sure it is level |
Level with Front Leveling Feet, or move the autoclave to a level surface |