Midmark M9 & M11 Error C570: POWER UP MODE
Display/Symptom |
Cause |
What to Check |
Solution |
C570: POWER UP MODE
STEAM TEMP OVERLIM
UNPLUG / REPLUG UNIT
Steam temperature exceeds 284°F (140°C) during POWER UP mode |
Power interruption due to a storm or power fluctuation. |
Allow Autoclave to cool |
Unplug unit for 60 seconds, then plug unit into a dedicated, properly rated circuit |
|
The operator may have unplugged the unit when a previous error code was displayed |
Check previous error codes. Refer to: Service Diagnostics (Recall Errors) |
Find the error code(s) in these troubleshooting charts |
|
Temperature sensor malfunctioning |
Be sure there is nothing touching the temperature sensor. (Example: trays, pouches, cassettes, etc) |
Reposition items so nothing touches the temperature sensor |
|
|
Check the temperature sensor probe for corrosion or residue. |
Clean the temperature sensor with Mid-Clean and distilled water. Click Here to get Mid-Clean (Part #MIC143-6907) |
|
|
Perform Temperature Sensor Testing |
Follow instructions in test procedures |
|
Sterilizer is not level |
Check surface where autoclave is located to make sure it is level |
Level with Front Leveling Feet, or move the autoclave to a level surface |