ERROR CODE C571: SELECT

Free Autoclave Troubleshooting Guides






Midmark M9 & M11 Error C571: SELECT


Display/Symptom Cause What to Check Solution
C571: SELECT
STEAM TEMP OVERLIM
UNPLUG / REPLUG UNIT

Steam temperature exceeds 284°F (140°C) during SELECT mode
Temperature sensor malfunctioning Be sure there is nothing touching the temperature sensor. (Example: trays, pouches, cassettes, etc.) Reposition items so nothing touches the temperature sensor
Check the temperature sensor probe for corrosion or residue. Clean the temperature sensor with Mid-Clean and distilled water. Click Here to get Mid-Clean (Part #MIC143-6907)
Perform Temperature Sensor Test Follow instructions in test procedures
Sterilizer is not level Check surface where autoclave is located to make sure it is level Level with Front Leveling Feet, or move the autoclave to a level surface