Midmark M9 & M11 Error C575: VENT MODE
Display/Symptom |
Cause |
What to Check |
Solution |
C575: VENT MODE
STEAM TEMP OVERLIM
ITEMS NOT STERILE
UNPLUG / REPLUG UNIT
Steam temperature exceeds 284°F (140°C) during STERILIZE Mode |
Pressure leaking from chamber |
Check for pressure leaks. Click Here To See "How To Find Pressure Leaks" |
Repair or replace malfunctioning component |
|
Temperature sensor malfunctioning |
Be sure there is nothing touching the temperature sensor. (Example: trays, pouches, cassettes, etc) |
Reposition items so nothing touches the temperature sensor |
|
|
Check the temperature sensor probe for corrosion or residue. |
Clean the temperature sensor with Mid-Clean and distilled water. Click Here to get Mid-Clean (Part #MIC143-6907) |
|
|
Perform Temperature Sensor Testing |
Follow instructions in test procedures |
|
Sterilizer is not Level |
Check surface where autoclave is located to make sure it is level |
Level with Front Leveling Feet, or move the autoclave to a level surface |