Midmark M9 & M11 C634: STERILIZE MODE
Display/Symptom |
Cause |
What to Check |
Solution |
C634: STERILIZE MODE
PRESSURE LOW
ITEMS NOT STERILE
UNPLUG / REPLUG UNIT
| Pressure leaking from chamber |
Check for pressure leaks. Click Here To See "How To Find Pressure Leaks" |
Repair or replace malfunctioning component |
|
Temperature sensor malfunctioning |
Be sure there is nothing touching the temperature sensor. (Example: trays, pouches, cassettes, etc) |
Reposition items so nothing touches the temperature sensor |
|
|
Check the temperature sensor probe for corrosion or residue |
Clean the temperature sensor with Mid-Clean™ Part #MIC143-6907 and distilled water |
|
|
Perform Temperature Sensor Testing |
Follow instructions in test procedures |
|
Sterilizer is not level |
Check surface where autoclave is located to make sure it is level |
Level with From Leveling Feet, or move the autoclave to a level surface |