Main PC Board
Pressure Transducer Test
This Test Checks For Proper Supply Voltage To The Pressure Transducer
The Pressure Transducer is located on the Newer Midmark Autoclave as shown in the picture below
(1) Set Meter To 20 VDC
(2)Place Probes From Meter To Pins 1 & 3 As Shown
Note: Pin Position Is Referenced Top To Bottom (Top Pin = 1 Bottom Pin = 4)
Meter Reading |
Action Required |
Acceptable Range 4.0 to 6.0 VDC |
Main PC Board Okay |
Any Reading Outside Of Acceptable Range |
Replace Main PC Board |