Chamber Over Filling

This Troubleshooting Guide is applicable for the following Midmark Autoclave Model Numbers: M9-001 thru -019, M9D-001 thru -019, M11-001 thru -019, M11D-001 thru -019

Notice Red Led Display


If Your Midmark has a Red LED Display as shown above, then this is the right guide for you

Error Code "C001"
(Low Water Error)
Applies to M9D & M11D Autoclaves Only


Display/Symptom Cause What to Check Solution
Chamber Over Fills With Water Temperature Sensor Is Malfunctioning Disconnect Temperature Sensor; Then Check For Proper Operation Replace Temperature Sensor