Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
Midmark Troubleshooting
›
M9 & M11 ERROR CODES-OLDER UNITS
›
Error Code E001
› E001 - DRYING CYCLE
E001 - DRYING CYCLE
You Will Need To Know When You Are Getting The E001 Error Code (As Soon As You Turn It On, or During the Heat up /Sterilization Cycle, or During The Drying Cycle) Before Proceeding
Midmark M9 & M11 Error Code E001: DRYING CYCLE
Display/Symptom
Cause
What to Check
Solution
E001: DRYING CYCLE
E001 Appears During Drying Cycle
Overheat Thermostat Activated - Caused By High Input Voltage
Check facility input voltage. Voltage should not exceed Electrical Requirements
Adjust the Dry Cycle Time Dip Switches (
Click Here
for Instructions)
Check Resistance of Heating Element (
Click Here
and Scroll Down the Page For Instructions for M9 units, Or
Click Here
and Scroll Down The Page For Instructions on M11 Autoclaves. If Resistance is out of Range, Replace Heating Element
Perform Continuity Check On Overheat Thermostat (Normally Closed) - Replace if Defective