E001 - HEAT UP / STERILIZATION MODE

You Will Need To Know When You Are Getting The E001 Error Code (As Soon As You Turn It On, or During the Heat up /Sterilization Cycle, or During The Drying Cycle) Before Proceeding

Midmark M9 & M11 Error Code E001: STERILIZATION CYCLE


Display/Symptom Cause What to Check Solution
E001: STERILIZATION CYCLE E001 Appears During Sterilization Cycle Bad Electrcial Connection - or

Overheat Thermostat Activated Because of:
  1. Pressure Leak
  2. Not Filling Properly
  3. Faulty Overheat Thermostat (Stuck Open)
Check sterilizer and facility for proper electrical connections.Tighten / replace loose connections
  1. Click Here for Instructions on How To Locate Pressure Leaks
  2. Click Here For Instructions on How To Test the Water Level Sensor
  3. Click Here For Instructions on How To Test The Overheat Thermostat