Midmark M9 & M11 Error Code E001: STERILIZATION CYCLE
Display/Symptom |
Cause |
What to Check |
Solution |
E001: STERILIZATION CYCLE |
E001 Appears During Sterilization Cycle
| Bad Electrcial Connection - or
Overheat Thermostat Activated Because of:
- Pressure Leak
- Not Filling Properly
- Faulty Overheat Thermostat (Stuck Open)
|
Check sterilizer and facility for proper electrical connections.Tighten / replace loose connections
- Click Here for Instructions on How To Locate Pressure Leaks
- Click Here For Instructions on How To Test the Water Level Sensor
- Click Here For Instructions on How To Test The Overheat Thermostat
|