E001 - START UP MODE

You Will Need To Know When You Are Getting The E001 Error Code (As Soon As You Turn It On, or During the Heat up /Sterilization Cycle, or During The Drying Cycle) Before Proceeding

Midmark M9 & M11 Error Code E001: START UP MODE


Display/Symptom Cause What to Check Solution
E001: POWER ON MODE Power Interruption -Autoclave Lost Power (Power interruption could be due to an electrical storm, brown out, etc.)
  • Autoclave Was Unplugged From the Wall Outlet & Plugged Back In
  • GFI Outlet Was Tripped
  • Circuit Breaker Was Tripped
  • Power Cord is defective
Check power cord for breaks - Replace if defective Press the STOP Button - The Autoclave will Reset. Then simply select a cycle as normal