Midmark M9 & M11 Error Code E001: START UP MODE
Display/Symptom |
Cause |
What to Check |
Solution |
E001: POWER ON MODE |
Power Interruption -Autoclave Lost Power (Power interruption could be due to an electrical storm, brown out, etc.)
- Autoclave Was Unplugged From the Wall Outlet & Plugged Back In
- GFI Outlet Was Tripped
- Circuit Breaker Was Tripped
- Power Cord is defective
|
Check power cord for breaks - Replace if defective |
Press the STOP Button - The Autoclave will Reset. Then simply select a cycle as normal |