E006

This Troubleshooting Guide is applicable for the following Midmark Autoclave Model Numbers: M9-001 thru -019, M9D-001 thru -019, M11-001 thru -019, M11D-001 thru -019

Notice Red Led Display


If Your Midmark has a Red LED Display as shown above, then this is the right guide for you

Midmark M9 & M11 Error Code E006


Display/Symptom Cause What to Check Solution
Error code 6
[Display Temperature Exceeded 277°F (136°C)]
  1. Slow Pressure Leak
  2. Sterilizer Potentiometers out of Calibration and/or Dirty Temperature Sensor
  3. Temperature Sensor Malfunctioning
  1. Check For Pressure Leaks (Click Here for Instructions on How To Locate Pressure Leaks)
  2. Clean Temperature Sensor; then Check Calibration using Our Midmark Calibration Kit For Field Sevice
  3. Check Calibration using our Midmark Calibration Kit For Field Service
  1. Replace Malfunctioning Component at Pressure Leak
  2. Perform Temperature / Pressure Potentiometer Adjustment
  3. Replace Temperature Sensor Assembly