Error Code E010 (Watchdog Timer Reset Error)

Midmark M9 & M11 Error Code E010
(Watchdog Timer Reset Error)


Display/Symptom Cause What to Check Solution
Error Code E010 (Watchdog Timer Reset Error)

Sterilizer does not complete cycle
  1. Control PC Board has had a software / hardware failure
  1. Disconnect power; then check for disconnected wiring in sterilizer or facility. Connect power & run additional cycle
  1. If error code persists, replace control PC board

allclaveparts.com widget logo