Error Code E010 (Watchdog Timer Reset Error)

This Troubleshooting Guide is applicable for the following Midmark Autoclave Model Numbers: M9-001 thru -019, M9D-001 thru -019, M11-001 thru -019, M11D-001 thru -019

Notice Red Led Display


If Your Midmark has a Red LED Display as shown above, then this is the right guide for you

Midmark M9 & M11 Error Code E010
(Watchdog Timer Reset Error)


Display/Symptom Cause What to Check Solution
Error Code E010 (Watchdog Timer Reset Error)

Sterilizer does not complete cycle
  1. Control PC Board has had a software / hardware failure
  1. Disconnect power; then check for disconnected wiring in sterilizer or facility. Connect power & run additional cycle
  1. If error code persists, replace control PC board