E011 - (Software Interrupt Error)

Free Autoclave Troubleshooting Guides



Midmark M9 & M11 Error Code E011
(Software Interrupt Error)


Display/Symptom Cause What to Check Solution
Error Code E011 (Software Interrupt Error)

Sterilizer does not complete cycle
  1. Control PC board has a Software / Hardware malfunction
  1. Disconnect power for approx. 10 seconds; then connect power & run additional cycle
  1. If error code persists, replace control PC board

allclaveparts.com widget logo