Skips Fill Cycle

This Troubleshooting Guide is applicable for the following Midmark Autoclave Model Numbers: M9-001 thru -019, M9D-001 thru -019, M11-001 thru -019, M11D-001 thru -019

Notice Red Led Display


If Your Midmark has a Red LED Display as shown above, then this is the right guide for you

Autoclave Skips Fill Cycle
Goes Directly To Heat Up Mode When Start Button Depressed


Display/Symptom Cause What to Check Solution
Skips Fill Cycle
  1. Water Level Sensor is Wet from Previous Cycle
  2. Water Level Sensor is Dirty
  3. Water Level Sensor is Defective
  1. Test Water Level Sensor
  1. Dry Water Level Sensor with Cloth or Allow To Air Dry
  2. Clean Face of Water Level Sensor With Abrasive (Scotch-Brite) Pad & Dry With Cloth
  3. Replace Water Level Sensor