Chamber Overfills With Water
Display/Symptom |
Cause |
What to Check |
Solution |
Water Pouring Out When Door is Opened |
- Water Level Sensor Inside Chamber is Malfunctioning or Dirty
- Loose, Broken, or Dirty Connection Between Chamber Water Level Sensor and Control PC Board
- Sterilizer is Not Level
- Ground Screw on PC Board Missing
- Temperature Sensor Malfunctioning
- Fill Solenoid Malfunctioning or Stuck
- Control PC Board Malfunctioning
|
- Unplug Sensor from PC Board & Jumper Connection on Board. Check for Proper Operation
- Perform Continuity Check Between Connector and Chanmber Water Level Sensor
- Make Sure Autoclave Is Level
- Check PC Board Ground Screw (Located on Lower Right Hand Corner on Board
- Unplug Temperature Sensor from PC Board Then Check for Proper Operation
- Check Resistance of Fill Solenoid
- Check all Probable Causes Listed Above
|
- Replace / Clean Water Level Sensor
- Replace / Clean Broken Wires or Connections
- Adjust Height of Individual Foot Levelers, or Shim Up Sterilizer to Make It Level
- Replace Ground Screw
- Replace Temperature Sensor
- Replace / Clean Fill Solenoid Valve
- Replace Control PC Board
|