Testing The Midmark M11 Pulse Solenoid

This Troubleshooting Guide is applicable for the following Midmark Autoclave Model Numbers: M9-001 thru -019, M9D-001 thru -019, M11-001 thru -019, M11D-001 thru -019

Notice Red Led Display


If Your Midmark has a Red LED Display as shown above, then this is the right guide for you

WARNING: The failure of the Pulse Solenoid can be fatal to your Control PC Board!

How to Test The Midmark M11 Pulse Solenoid

 

This is a simple test, but you will need an Ohm Meter to complete it

Perform a resistance check. First disconnect the two wires to the coil, then perform the resistance check, looking for a short circuit.  With your meter set for 200 ohms (make sure your batter in the meter is fresh), place a probe on each of the two terminals and look at the reading

The resistance of the coil should read as follows:   Pulse Solenoid (Part #MIS079- 817) ~11.6 Ohms

If the resistance reading of the Coil is near zero ohms, the Coil is shorted and needs to be replaced