Wet Instruments

This Troubleshooting Guide is applicable for the following Midmark Autoclave Model Numbers: M9-001 thru -019, M9D-001 thru -019, M11-001 thru -019, M11D-001 thru -019

Notice Red Led Display


If Your Midmark has a Red LED Display as shown above, then this is the right guide for you

Instruments Not Drying


Display/Symptom Cause What to Check Solution
Instruments Are Wet After Drying Cycle Is Complete
  1. Door Is Not Opening At End Of Vent Cycle
  2. Sterilizer Is Not Level
  3. Filter Is Plugged
  1. See Problem: Door Does Not Open After Vent Cycle By Clicking Here
  2. Check Sterilizer For Level Installation
  3. Check To Make Sure There Is No Water Is Left In Chamber After Venting
  1. Click Here For Further Instructions
  2. Re-install Sterilizer On A Level Surface, Adjust Height Of Individual Foot Levelers, or Shim Up Sterilizer
  3. Replace / Clean Filter