Chamber Does Not Vent Properly
Display/Symptom |
Cause |
What to Check |
Solution |
Door opens after a longer than normal venting period on M9 and / or M11 Autoclaves
OPEN THE DOOR Lamp Illuminates After Longer Than Normal Venting Period on M9D and / or M11D Autoclaves |
- Drain, Filter, or Vent Tubing Plugged
- Vent Solenoid Malfunctioning or Stuck.
|
- Remove Filter and Tubing and Inspect
- Check Resistance of Vent Solenoid Coil
|
- Clean, Drain, and Flush Tubing. Replace / Clean Filter
- Replace / Clean Vent Solenoid Valve
|