Simple Bellows Test
Is the bellows faulty? When the temperature range stays between 210° & 212° it is indicative of a faulty bellows
To confirm it is the bellows:
- Remove the cabinet by removing the 3 screws on each side of the autoclave. Then simply lift it off
- Now remove the lid off of the water reservoir so you can see the pipe inside the water reservoir coming from the bellows
- Run a cycle
- When it gets to 210° look at the end of the pipe inside the reservoir
- If you see steam coming out of the pipe, then the bellows is faulty and will need to be replaced
We offer two types of replacement bellows.
The Diaphragm Style RCB089-787
and
The Bellows Style RCB100-849.
The Bellows Style, while more expensive, is designed to last longer.
Replacing the Bellows is also simple
- Using a 1/2" open-end wrench, remove the nuts from each side of the bellows (disconnecting the pipes from the bellows assembly)
- Install the new bellows with the Flow Arrow Pointing up to the water reservoir, reconnecting the pipes with the open end wrench
- Run a cycle and make sure you do not have any leaks at the connection points
You are done...and you have saved the cost of a service call!