Troubleshooting The Midmark M9 or M11 Sterilizer (Models M9, M9D, M11, M11D)
|
M9/M11 Door Opening Test
NOTE: This test only applies to units with the newer style PC Boards
- Fill chamber with water. (To do this, start any cycle. After chamber has filled with water, the Sterilizing LED will illuminate)
- Disconnect Power Cord
- Open Door
- Press & Hold STOP & POUCHES buttons while plugging in power cord
- Press & Release all buttons from left to right across display panel
- Close door. (This will activate all relays in the sterilizer in sequence. Pulse solenoid should automatically open door when test cycle is complete)
- To rerun test, repeat steps 2 thru 6
- To verify proper operation, run a complete cycle when test is complete
|