Panel 1: Areas To Check For Pressure Leaks

Free Autoclave Troubleshooting Guides













AllClaveParts.com

Troubleshooting The Midmark M9 or M11 Sterilizer
(Models M9, M9D, M11, M11D)

Areas To Check For Pressure Leaks


mr23a (51K)
Component Check Correction
Door Gasket (1) Check for water leaking around door (See 2) Replace Door Gasket
Vent Solenoid (3) Check For Water Leaking thru condensing coil (See 4) Replace Vent Solenoid
Fill Solenoid (5) Check for water leaking back into reservoir thru fill line, raising reservoir water level (See 6) Replace Fill Solenoid
Bellows (7) Chck for excessive steam from Bellows line in reservoir. NOTE: A light hiss of steam is not uncommon (See 8) Replace Bellows
Pressure Relief Valve (9) Check for water leaking beneath back of sterilizer. NOTE: Check actual pressure before replacing pressure relief valve (see 10) Replace Pressure Relief Valve
Pressure Sensor Hose (11) Check for steam leak on main PC Board (see 12) Secure pressure sensor hose to PC Board with cable tie

allclaveparts.com widget logo