Right Hand Side Panel Removal/Installation


WARNING Always disconnect the power cord from the outlet before removing any of the sterilizer covers/panels or making any repairs to prevent the possibility of electrical shock. Also, drain the sterilizer to prevent spills during repairs. Failure to comply with these instructions could result in serious personal injury or death.


Right Hand Side Panel Removal/Installation
For The Midmark M9 or M11 Sterilizer
(Models M9, M9D, M11, M11D)

Removal

  1. Disconnect power cord from the outlet.
  2. Remove two screws (1) from top cover (2).
  3. Remove left screw (1) from steam block (2)
  4. Remove two screws (3) from top cover (4)
  5. Lift up the left hand corner of top cover (4), pull outward and down on the top edge of the left hand side panel (5), and remove left hand side panel from sterilizer.

mrs (39K)

Installation

  1. Insert two tabs of left hand side panel (5) into two slots of base (6).
  2. Lift up the left hand corner of top cover (4), raise the top edge of left hand side panel (5) into position, and secure using two screws (3).
  3. Install left screw (1) on steam block (2).
  4. Close sterilizer door.