Before Replacing Any Electrical Or Mechanical Componet, Unplug the Autoclave From The Outlet. If Autoclave Must Be Powered On
For Any Repair Procedures, USE EXTREME CAUTION
NAPCO 8000DSE Autoclave Troubleshooting
Problem |
Procedure |
No Power
Temperature Light Out
Autoclave Does Not Heat |
- Check Plug, Outlet & Circuit Breaker
- Check Fuses
- Check Switches
- Check Timer
|
Steam Pressure Is Lost Water Boils Our Before Sterilization Is Achieved |
- Check Water Level At Start Of Each Cycle
- Check For Water Leaks In Tubing & Fittings
- Check Purge Systems
- Re-calibration Of Sensor Is Advised For Operation At High Elevations
|
Temperature Light Fails To Illuminate, Autoclave Heats Normally |
Check Temperature Light |
Time Light Fails To Illuminate |
Check Timer Light |
Autoclave Does Not Heat
Time Light Illuminated - Temperature Light Out
|
- Check Low Water Cut Off Switch
- Check Dryer Switch
- Check Autoclave Temperature Control
|
Autoclave Does Not Heat
Temperature Light Illuminated
|
Check Heating Element |
Autoclaving Temperature Control Cannot Be Achieved Or Maintained
Pressure Gauge Has A Pressure Reading
|
- Check Autoclave Temperature Control
- Check Calibration Adjustments
|
Autoclave Temperature Control Cannot Reach Maximum Or Minimum Limits
(121°C to 132°C)( +3°C) By Turning Knob
|
Check Calibration Adjustment For Temperature Control |
Temperature Fluctuations of 4°C Or More Detected On Temperature Gauge At Preset Autoclaving Temperature |
- Check Gauges
- Check Autoclave Temperature Control
Note: 10 Minutes Should Be Allowed Between Shutdown & Restart
|
Timer Does Not Function Properly
Time Light Is Illuminated
|
Check Timer |
Drying Element Does Not Heat
Time Light Illuminated
Dryer Light Illuminated
Temperature Light Out
|
- Check Dryer Temperature Control
- Check Drying Element
|
Fan Motor Does Not Run
8000DSE Only |
Autoclave Must Be In Slow Exhaust Function |
Autoclave Runs Out Of Water
Autoclave Fails To Shut Off At Excessive Temperature Resulting In Heat Damage |
Check Low Water Cut Off Switch |
Autoclave Temperature Control Failure (Excessive Heating)
Pressure Is Relieved At Safety Pressure Of 35psi ( + 3psi
|
Check Autoclave Temperature Control |