Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
NAPCO Troubleshooting
›
NAPCO 704-9000D TROUBLESHOOTING
› NAPCO 704-9000D CONTROL ADJUSTMENTS
NAPCO 704-9000D CONTROL ADJUSTMENTS
Yes, You Can Adjust the Temperature Control on Your NAPCO 704-9000D Autoclave
Just Follow The Directions Below
NAPCO 704-9000D Control Adjustments
Temp-Autoclave Control
The Temp-Autoclave Control may require adjustment. The function of this control is to maintain a maximum stabilized temperature of 132°C (270°F). The control requires calibration to assure that it is at the proper setting
To calibrate the control the following procedure should be followed:
Check the water level and fill to gauge level
Turn Function Switch to AUTOCLAVE ONLY. Turn Power Switch ON & Then Start The Timer
Heat Unit to 132°C (270°F) by Adjusting the Temp-Autoclave Control Shaft
To gain access to the Temperature Control Shaft, remove the knob by loosening the two set screws. The slotted shaft can now be seen
Turn the shaft counterclockwise to raise the temperature. Turn clockwise to lower the temperature
Watch Temperature Gauge on door and adjust by 1/2 turn increments until temperature is reached
Allow the unit to stablize. Watch the Temperature Gauge. The Temperature Light will flash off and on at intervals holding the temperature within tolerance
Adjust as necessary during the stablization time. Temperature Light should flash on each time there is a 1° drop in temperature. When the unit has flashed several times the unit is stablized
Place knob on temperature control shaft with indicator at position 7 and tighten both set screws
Check operation of shaft and knob by turning to position 1 and back to position 7
Adjusting The Sensor
The sensor is set at the factory (near sea level) to close the Solenoid at 2°C above the boiling point of water. Since the boiling point of water will vary with altitude the Sensor may need adjustment when setting up the unit at higher elevations. Note the applicable temperature on the following chart..
BOILING TEMPERATURE OF WATER VS ALTITUDE
Altitude (Meters)
Temperature °C
Altitude (Feet)
Temperature °F
0
100
0
212
150
99.5
1000
210.1
300
98.9
2000
208.3
600
97.9
3000
206.5
900
96.9
4000
204.6
1200
95.9
5000
202.8
1500
94.8
6000
201.0
2000
92.9
7000
199.3
See Illustration Below
Set Temperature Autoclave Control on Position 4, turn Power Switch ON and activate Timer (Check Water Level In Chamber Before Turning the Autoclave On
Observe the temperature at which the Solenoid is activated. Listen for the first Click of the solenoid during warm-up
If Sensor requires adjusment to 2°C above boiling point, locate adjustment screw at the end of the Sensor
CAUTION:
the terminal next to the adjustment screw is Electrically Hot. Use insulated screwdrive to make adjustment
Turn the Power Switch OFF and allow the unit to cool for 10 minutes
Repeat Step 1 above and observe the Temperature Gauge
When the predetermined Temperature is obtained (2°C above local boiling point of water) turn Sensor adjusting screw counterclockwise until the Solenoid is activiated
The autoclave should then be cooled and reactivated to activate the Solenoid again. This will confirm that the Sensor adjustment is now correct
Click Here
For A Larger Image
Click Here
For A Larger Image