The Following Procedures Are For The NAPCO 9000D Autoclave Only



How To Test The NAPCO 9000D Autoclave Timer



  1. Turn Power Switch To "ON" Position. Set Timer For 20 Minutes & Observe The Timer In Operation
  2. If The Rotation Speed Of The Timer Is Not Accurate, It Is Defective & Should Be Replaced
  3. If The Time Light Is "ON" But The Timer Does Not Operate, Test With Voltmeter Across Terminals R-1 & L-3 (See Wiring Diagram)
    If No Voltage Here, Make Sure All The Connections Are Tight And There Is Power To The Timer
    Retest & If Still Defective, Replace Timer