Symptom |
Cause |
Solution |
NO CONTROL OF PRESSURE/TEMPERATURE |
A) OVERTEMP SWITCH |
A) RECALIBRATE FROM COLD START. |
|
B) THERMISTOR ASSEMBLY |
B) RESISTANCE SHOULD BE 1 MEG W@ 77°F +20%. |
|
C) SOLID STATE CONTROLLER |
C) MEASURE VOLTAGE FROM COMMON TERMINAL OF POWER SWITCH NO.2 TO COMMON TERMINAL OF FUNCTION SWITCH. IF 120 VOLTS, BOARD IS OPEN. IF 0 VAC, BOARD IS GOOD.
|
|
D) CENTER HEATING ELEMENT |
D) POSSIBILITY OF SHORT IN HOT SIDE OF LINE. TRACE OUT HOT
SIDE OF LINE FROM POWER CORD TO SWITCHES FOR A SHORT.
CHECK CENTER ELEMENT FIRST. |
WILL NOT BUILD UP TO STERILIZING PRESSURE/TEMPERATURE |
A) DOOR |
A) CHECK DOOR TO MAKE SURE IT IS IN THE LOCKED POSITION AND NOT LEAKING. |
|
B) DOOR GASKET |
B) REPLACE DOOR GASKET IF LEAKING OCM Door Gasket OCR Door Gasket OCR Plus Door Gasket |
|
C) DOOR SWITCH (OCR+ ONLY) |
C) CHECK SWITCH POSITION AND READJUST IF NECESSARY. |
|
D) UNIT LEVEL |
D) USE LEVEL AND ADJUST FEET IF NECESSARY. |
|
E) AMOUNT OF WATER ADMITTED DURING FILL CYCLE |
E) WHEN PROPERLY FILLED, WATER SHOULD COVER FILL PLATE. |
|
F) POTENTIOMETER ON CONTROLLER |
F) TURN CLOCKWISE TO CALL FOR MORE HEAT. |
|
G) CENTER HEATING ELEMENT |
G) CHECK RESISTANCE OF ELEMENT, IT SHOULD BE:
~ 10W FOR OCM
~ 8W FOR OCR OR OCR+ |
|
H) OVERTEMP SWITCH |
H) RECALIBRATE FROM COLD START, OR REPLACE AND CALIBRATE. OCM OVERTEMP SWITCH OCR OVERTEMP SWITCH OCR Plus OVERTEMP SWITCH |
|
I) THERMISTOR |
I) CHECK RESISTANCE. SHOULD BE 1 MEG W@ 77°F +20% |
|
J) FRONT POTENTIOMETER |
J) CHECK RESISTANCE 20K W +10%. |
|
K) SOLID STATE CONTROLLER |
K) MEASURE VOLTAGE FROM TOP TERMINAL OF NO. 1 POWER SWITCH TO COMMON TERMINAL OF FUNCTION SWITCH. (IF 120 VAC BOARD IS PROBABLY OPEN; IF 0 VAC, BOARD IS PROBABLY GOOD). THIS IS PROVIDED THERE IS CONTINUITY THROUGH FUNCTION SWITCH, OT SWITCH AND HEATING ELEMENT. |
|
L) MAIN VALVE (FILL/VENT VALVE) |
L) CHECK TEFLON SEATS FOR SCORING OR TRASH LODGED AROUND SEATS. IF SCORED, REPLACE. IF TRASH OR DEBRIS, CLEAN AND RE-ASSEMBLE. |
|
M) FUNCTION SWITCH |
M) CHECK FOR ACTIVATION OF SWITCH BY MAIN VALVE CAM. ADJUST POSITION OF SWITCH ARM SO IT IS DEPRESSED AND SWITCH IS CLOSED IN THE STERILIZER MODE. |
|
N) ROCKER ARM |
N) VISUALLY CHECK FOR ROCKER ARM DEPRESSING PUSH ROD IN STERILIZE MODE. CHECK FOR CLEARANCE AND ADJUST (FLATTEN) CAM SHAFT IF NECESSARY. |
THERMOMETER READS MORE THAN 4°F BELOW REQUIRED TEMPERATURE AT GIVEN PRESSURE |
A) BELLOWS |
A) REPLACE BELLOWS, IF NECESSARY OCM BELLOWS OCR BELLOWS OCR Plus BELLOWS |
|
B) TEMPERATURE GAUGE
Article On Temperature Gauges
|
B) REPLACE TEMPERATURE GAUGE, IF NECESSARY OCM TEMPERATURE GAUGE OCR TEMPERATURE GAUGE OCR Plus TEMPERATURE GAUGE |
|
C) PRESSURE GAUGE |
C) REPLACE PRESSURE GAUGE, IF NECESSARY OCM PRESSURE GAUGE OCR PRESSURE GAUGE OCR Plus PRESSURE GAUGE |
|
D) AIR VALVE HOUSING |
D) CLEAN AIR VALVE HOUSING AND/OR REPLACE. |
THERMOMETER READS MORE THAN 4°F ABOVE REQUIRED TEMPERATURE AT GIVEN PRESSURE |
A) TEMPERATURE GAUGE
Article On Temperature Gauges
|
A) REPLACE TEMPERATURE GAUGE, IF NECESSARY. OCM TEMPERATURE GAUGE OCR TEMPERATURE GAUGE OCR Plus TEMPERATURE GAUGE |
|
B) PRESSURE GAUGE |
B) REPLACE PRESSURE GAUGE, IF NECESSARY. OCM PRESSURE GAUGE OCR PRESSURE GAUGE OCR Plus PRESSURE GAUGE |
PILOT LIGHT BULBS SLOW TO LIGHT WHEN AUTOCLAVE IS TURNED ON |
A) LINE VOLTAGE |
A) CHECK LINE VOLTAGE. |
|
B) BULB |
B) REPLACE BULB OCM Bulb OCR Bulb OCR Plus Bulb |
TIMER BELL WILL NOT RING AT END OF TIMING CYCLE |
A) TIMER |
A) REPLACE TIMER, IF NECESSARY OCM BELL TIMER OCR BELL TIMER OCR Plus BELL TIMER |
|
B) TIMER BELL OFF-CENTER |
B) LOOSEN NUT ON BACKSIDE OF BELL AND RECENTER. |
|
C) TIMER POSITION |
C) LOOSEN NUT ON TIMER SHAFT, ROTATE TIMER UNTIL TOP MARK IS AT 12:00 POSITION. |