OCM,OCR & OCR+ Troubleshooting Guide

Free Autoclave Troubleshooting Guides



AllClaveParts.com

ocm_acp (27K)

Troubleshooting Guide For Maintaining & Repairing Your Pelton Crane
OCM & OCR Sterilizer

Troubleshooting



Symptom Cause Solution
NO CONTROL OF PRESSURE/TEMPERATURE A) OVERTEMP SWITCH A) RECALIBRATE FROM COLD START.

B) THERMISTOR ASSEMBLY B) RESISTANCE SHOULD BE 1 MEG W@ 77°F +20%.

C) SOLID STATE CONTROLLER C) MEASURE VOLTAGE FROM COMMON TERMINAL OF POWER SWITCH NO.2 TO COMMON TERMINAL OF FUNCTION SWITCH. IF 120 VOLTS, BOARD IS OPEN. IF 0 VAC, BOARD IS GOOD.


D) CENTER HEATING ELEMENT D) POSSIBILITY OF SHORT IN HOT SIDE OF LINE. TRACE OUT HOT SIDE OF LINE FROM POWER CORD TO SWITCHES FOR A SHORT. CHECK CENTER ELEMENT FIRST.
WILL NOT BUILD UP TO STERILIZING PRESSURE/TEMPERATURE A) DOOR A) CHECK DOOR TO MAKE SURE IT IS IN THE LOCKED POSITION AND NOT LEAKING.

B) DOOR GASKET B) REPLACE DOOR GASKET IF LEAKING
OCM Door Gasket
OCR Door Gasket
OCR Plus Door Gasket

C) DOOR SWITCH (OCR+ ONLY) C) CHECK SWITCH POSITION AND READJUST IF NECESSARY.

D) UNIT LEVEL D) USE LEVEL AND ADJUST FEET IF NECESSARY.

E) AMOUNT OF WATER ADMITTED DURING FILL CYCLE E) WHEN PROPERLY FILLED, WATER SHOULD COVER FILL PLATE.

F) POTENTIOMETER ON CONTROLLER F) TURN CLOCKWISE TO CALL FOR MORE HEAT.

G) CENTER HEATING ELEMENT G) CHECK RESISTANCE OF ELEMENT, IT SHOULD BE:
• ~ 10W FOR OCM
• ~ 8W FOR OCR OR OCR+

H) OVERTEMP SWITCH H) RECALIBRATE FROM COLD START, OR REPLACE AND CALIBRATE.
OCM OVERTEMP SWITCH
OCR OVERTEMP SWITCH
OCR Plus OVERTEMP SWITCH

I) THERMISTOR I) CHECK RESISTANCE. SHOULD BE 1 MEG W@ 77°F +20%

J) FRONT POTENTIOMETER J) CHECK RESISTANCE 20K W +10%.

K) SOLID STATE CONTROLLER K) MEASURE VOLTAGE FROM TOP TERMINAL OF NO. 1 POWER SWITCH TO COMMON TERMINAL OF FUNCTION SWITCH. (IF 120 VAC BOARD IS PROBABLY OPEN; IF 0 VAC, BOARD IS PROBABLY GOOD). THIS IS PROVIDED THERE IS CONTINUITY THROUGH FUNCTION SWITCH, OT SWITCH AND HEATING ELEMENT.

L) MAIN VALVE (FILL/VENT VALVE) L) CHECK TEFLON SEATS FOR SCORING OR TRASH LODGED AROUND SEATS. IF SCORED, REPLACE. IF TRASH OR DEBRIS, CLEAN AND RE-ASSEMBLE.

M) FUNCTION SWITCH M) CHECK FOR ACTIVATION OF SWITCH BY MAIN VALVE CAM. ADJUST POSITION OF SWITCH ARM SO IT IS DEPRESSED AND SWITCH IS CLOSED IN THE STERILIZER MODE.

N) ROCKER ARM N) VISUALLY CHECK FOR ROCKER ARM DEPRESSING PUSH ROD IN STERILIZE MODE. CHECK FOR CLEARANCE AND ADJUST (FLATTEN) CAM SHAFT IF NECESSARY.
THERMOMETER READS MORE THAN 4°F BELOW REQUIRED TEMPERATURE AT GIVEN PRESSURE A) BELLOWS A) REPLACE BELLOWS, IF NECESSARY
OCM BELLOWS
OCR BELLOWS
OCR Plus BELLOWS

B) TEMPERATURE GAUGE
Article On Temperature Gauges
B) REPLACE TEMPERATURE GAUGE, IF NECESSARY
OCM TEMPERATURE GAUGE
OCR TEMPERATURE GAUGE
OCR Plus TEMPERATURE GAUGE

C) PRESSURE GAUGE C) REPLACE PRESSURE GAUGE, IF NECESSARY
OCM PRESSURE GAUGE
OCR PRESSURE GAUGE
OCR Plus PRESSURE GAUGE

D) AIR VALVE HOUSING D) CLEAN AIR VALVE HOUSING AND/OR REPLACE.
THERMOMETER READS MORE THAN 4°F ABOVE REQUIRED TEMPERATURE AT GIVEN PRESSURE A) TEMPERATURE GAUGE
Article On Temperature Gauges
A) REPLACE TEMPERATURE GAUGE, IF NECESSARY.
OCM TEMPERATURE GAUGE
OCR TEMPERATURE GAUGE
OCR Plus TEMPERATURE GAUGE

B) PRESSURE GAUGE B) REPLACE PRESSURE GAUGE, IF NECESSARY.
OCM PRESSURE GAUGE
OCR PRESSURE GAUGE
OCR Plus PRESSURE GAUGE
PILOT LIGHT BULBS SLOW TO LIGHT WHEN AUTOCLAVE IS TURNED ON A) LINE VOLTAGE A) CHECK LINE VOLTAGE.

B) BULB B) REPLACE BULB
OCM Bulb
OCR Bulb
OCR Plus Bulb
TIMER BELL WILL NOT RING AT END OF TIMING CYCLE A) TIMER A) REPLACE TIMER, IF NECESSARY
OCM BELL TIMER
OCR BELL TIMER
OCR Plus BELL TIMER

B) TIMER BELL OFF-CENTER B) LOOSEN NUT ON BACKSIDE OF BELL AND RECENTER.

C) TIMER POSITION C) LOOSEN NUT ON TIMER SHAFT, ROTATE TIMER UNTIL TOP MARK IS AT 12:00 POSITION.